AirShields TI 100 Infant Incubator
Condition REFURBISEHD
FEATURES
- Intended for transport of high risk, premature, low birth-weight, or critically ill newborns.
- It provides a means to control air temperature, oxygen concentrations, and relative humidity.
- A double wall hood permits full visibility and provides an effective thermal barrier from the environment.
SPECIFICATIONS
Dimensions (with stand):
- Length: 40.25 in. (102 cm)
- Width: 22.25 in. (56.5 cm)
- Height: 46.75 in. (118.7 cm) maximum, 34.75 in. (88.3 cm) minimum
Power Requirements:
- AC: 110.120 VAC 50/60/400 Hz, sine or square wave 170/200 Watts and 225/270 Watts with batteries charging.
- DC: 12 VDC, 115 Watts
Internal Battery:
- Type: gel type, sealed and rechargeable.
- Voltage: 12 VDC nominal
- Quantity: 2
- Charging Time: two batteries take 20 minutes to recharge from fully discharged.
- Life expectancy: minimum 200 full charge/discharge cycles.
Chassis Leakage Current: 100µA or less
Alarms:
- High Temp: Actuates at displayed incubator temperature of 39º ± 0.5 ºC.
- Sensor: Actuates if High Temp alarm sensor is electrically open or shorted.
- Power Failure: Actuates if main AC power fails and no DC back-up power is present.
- Heater Temp: Actuates if heater temperature exceeds 77 ºC nominal.
- Low DC: Actuates if incubator DC voltage has fallen below 10.5 VDC nominal.
Temperature:
- Set Point Range: 21.5 ºC ± 1.0 ºC to 38 ºC
- Rise Time: 30 minutes
- Variability: 0.2 ºC
- Overshoot: 1.5 ºC maximum
- Uniformity: 1 ºC
- Correlation of Displayed Incubator to Actual Incubator Temperature at Temperature Equilibrium: 1 ºC
- Correlation of Displayed Incubator Temperature to Set Point at Temperature Equilibrium: ± 0.5 ºC
- Observation Light Output: 10-foot candles at mattress level